The available field-of-view for COSMIC multi-slits is .
It is suggested that the width of the multi-slit pattern be less than 4 arc-min,
preferably even less than this,
to ensure that the spectra have approximately the same central wavelength (and hence
spectral coverage) and to avoid problems with spatial distortion.
Since the multi-slit masks are made by a photographic reproduction
process, the width of each slit can be set by the user.
The COSMIC slitmask assignment program takes a list of offsets (arc-sec East and arc-sec North, where East and North are positive) of objects from the center of a field. This program, called ``cosmicslitmask'', was written by Mike Pahre based on notes supplied by Alan Dressler. It is a simple program, assuming a constant plate scale (11.9''/mm) for the prime focus, and ignoring the complications of spatial distortion, gnomic projections, etc. The help file for this program is in anonymous ftp on phobos:/users/map/COSMIC/README (or can be accessed from phobos itself as ftp/users/map/COSMIC/README).
The output of this program consists of a text file and a plot file. You should create two separate plot files--one which includes the object number above each slit, the other with only the slits and the identification label. The latter plot is the one that needs to go to the photographic lab for reduction.
Note the field identification label on the upper left corner of the plot. The position angle for the mask in terms of the COSMIC ``base position angle'' is given in this label. The limits on rotation for the COSMIC base position angle is 70 to 300 degrees. (MAP put a software check into COSMICslitmask)
Make sure to take to the mountain printed copies of all 3 of these files (one text and two plots) for each mask you design. You should also have a good image of the field with all of the objects identified.
The slit masks used by the COSMIC spectrograph are made by reproducing photographically at a reduction factor of 4 the output of the COSMIC slit mask assignment program. Thus the slits correspond to places where the film is clear (but there is still some transmission loss of course), and the remaining area must be very opaque. If the contrast in transmission between the slit and the dark area is not at least 104, then sky light may well leak in and contaminate the exposure at a noticable level. Thus the two critical concerns for the photographic lab are the exact reduction by a factor of 4 of the supplied image and making the mask area (except for the slits) as opaque as possible.
Commercial Graphics (681 S. Raymond Ave., Pasadena, CA 91105) is the vendor of choice. They will have to send the materials to their LA Office to be photographed. The instructions to the vendor are to:
reduce 4 times
make a negative
reverse reading
Four masks are reproduced (reduced by a factor of 4) onto a sheet of 8.5 by 11 film.
To mount a photographic film slitmask onto a holder you will need:
light table; scissors; narrow black tape (for blocking out the mask
border); and a jeweler's screwdriver. All except for the black tape
can be provided by the day crew. The tape should be bought in
town; good examples are ``Letraline 476 3/16-inch black matt 3730''
or ``Letraline 96 1/4-inch black flex 2426''. Call around for a
graphic design store to find such tape. You may also use the black
photographic tape which is kept in the wooden desk just outside the control
room, on the observing floor.
Using scissors, trim the slitmask film to keep inch
outside the rectangular mask border. Also, diagonally cut off
the four corners of the film.
Take a mask holder and remove the cover (there should be four
screws holding it in position).
There are three notches cut into the outer ring of the holder.
One notch is closest to a short side of the rectangular
mask hole; place this notch facing towards your left.
Turn your slitmask upside down so that the label reads backwards
and lies in the upper right corner, and place on top of the
mask holder.
Place the cover on top, and lightly fix in place with the screws,
leaving it loose enough so that you can still move the film around.
Place the mask holder on the light table.
Slide the film around until the upper/lower mask border lines
up with the upper/lower edge of the rectangular cutout of the
mask holder.
This is a crucial step to do carefully, as it sets the rotation
angle of the slits in the mask.
When you are satisfied, tighten the four screws to fasten the
film in place.
Use the black tape to cover the four clear borders of the film
and the label. Also cover any imperfections in the black portion
of the emulsion.
Be sure to note on the tape the identification or name of the mask.
Use the blue box to move the desired aperture wheel position to
the install location. First push ``AP'' to select the
aperture motor, then push ``INC'' or ``DEC'' to move the wheel.
The position displayed by the blue box is the position of the
aperture wheel that is in the optical path; the position in the
installation location at that time is 2 positions less (modulo 6).
Reach over the blue motor housing mounted just above the aperture
wheel on COSMIC. Hold the holder
so the ``locating notch'' is pointing away from the blue
motor housing. Insert the holder and mask so that the 3 notches
all clear, then press down, and turn clockwise approximately
1/4 inch, after which you will hit a stop. The stop is defined
by knife blades hitting the beveled edge of the each of holder notches.
(CHECK ALL KNIFE BLADES ARE INSTALLED). For peace of mind, tape the
holder to the aperture wheel once the holder is installed.
Repeat this procedure using the blue box to move to each desired position
on the aperture wheel
for each mask you plan to use.
Be sure to close the velcro flap on the shroud prior to leaving
the prime focus cage.
Once you are back in the control room, cycle through
several different positions on the aperture wheel to
reset it properly.
For each mask you need to create a file containing your alignment objects.
This should
be done in the afternoon before observing, and the file should be in the
same directory that you are using for taking images. This can be done
easily by editing the tabular output of the
Take a direct image of each of the slit-masks you have just
installed. This images should be taken with the FILTER=CLEAR, GRISM=CLEAR,
and APERTURE=multislit of your choice.
Display the image of the desired slit-mask (with the [*,-*] orientation).
Once you have taken the images, use the IRAF script ``masksetup''
to mark the positions of the slits corresponding to the objects
in the ``mask.slits'' file. In IRAF, type ``masksetup''. You will be asked
for an output filename (typically ``mask'') and the input filename (typically
``mask.slits''). You will then be asked to press the ``b'' key
at the left and right edges of the slit corresponding to your first object,
followed by ``q'', then
the ``a'' key at the center of slit, followed by ``q''. This will be repeated
for each alignment slit. After the last slit, the calculated positions of the
objects will be marked on the screen, and an output
file called ``masks.ap'' will be produced which will be used during the night
for alignment.
There are 6 positions in the aperture wheel, position 1 of which must be left
clear for direct images. The 1.5 arc-sec long slit is usually left
in position 2, leaving 4 positions for slit-masks.
There are 6 holders for slit masks.
As you enter the prime focus cage of the 200-inch telescope, move
around counter-clockwise to provide easy access to the blue box
(the COSMIC local motor control console).
Assuming the zipper on the COSMIC dark shroud is in its usual place,
you also then will have easy access to a velcro-closed opening in the
shroud
at about waist level. Open this to get access to the aperture wheel.
The first time you try this you might want to unzip the zipper on the
COSMIC dark shroud, so that you can see the entire instrument more
easily. (Sometimes the setup crew leaves this shroud in a position
rotated from its correction orientation, so you might have to
remove the shroud anyway.)
You should write a tape at your home institution including the
tabular output files from cosmicslitmask.f. Dump these
files into your current COSMIC directory on the COSMIC Sparc
computer (i.e. where the data are being written).
cosmicslitmask.f program:
(1) remove the header line(s); (2) keep only 5-10 objects that
are good for slitmask alignment; and (3) reorder them so that the
objects go from left to right (in the plot of the slitmask).
A good object for slitmask alignment is relatively bright and
compact, so that an accurate centroid can be measured in a relatively
short imaging exposure. Call this file ``mask.slits''.